Skytron recalls Freedom 4FXS-60 Heavy Duty Monitor Mounts due to risk of frame separation from mounting column under weight. Affected units have specific serial numbers.
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Skytron is recalling its Freedom 4FXS-60 Heavy Duty Monitor Mounts because the mounting hardware may not hold the monitor and frame securely. This could cause the monitor frame to separate from the mounting column, potentially leading to injury.
The mounting hardware supplied with the 4FXS-60 bracket has inadequate thread engagement to support the monitor's weight. This could cause the frame to separate from the mounting column, potentially leading to injury if the monitor falls.
Users of the Skytron Freedom 4FXS-60 monitor mounts with the listed serial numbers are affected. People who mount heavy monitors or use the product without proper installation are at higher risk.
Check the serial number on the product. Affected units have the serial number 10841736116071 or other numbers listed in the recall notice.
Look for the serial number on the monitor mount to confirm if it's affected.
The recall states the mounting hardware may not hold the monitor securely, so it's not safe to use without addressing the issue.
Check the serial number on the product; affected units have the serial number 10841736116071 or other numbers listed in the recall notice.
The recall does not specify a remedy, so the user should check the serial number and contact Skytron for further instructions.
We’ve drafted a message you can send Skytron to request your refund or repair — edit it as you like.
Subject: Recall Z-1663-2025 — Skytron Skytron Hello, I own a Skytron that is covered by recall Z-1663-2025 from Skytron. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.