Digimed recalls portable X-ray systems with model DIOX-602 due to insufficient safety distance (SSD) of 18 cm. Affected units may not meet safety standards.
Some DIOX-602 units currently in use within the United States do not meet the minimum SSD requirement of 18 cm.
Digimed is recalling portable X-ray systems with model DIO:0X-602 because they don't meet the minimum safety distance requirement of 18 cm. This could lead to safety risks if used improperly.
The X-ray system requires a minimum safety distance (SSD) of 18 cm to prevent radiation exposure. If the SSD is smaller than this, it could increase the risk of radiation exposure to users or patients.
Healthcare facilities and medical professionals using Digimed DIOX-602 portable X-ray systems in the U.S. are most at risk due to potential safety issues.
Check for the model number DIOX-602 on the device. The product was sold in the U.S. with the UDI-DI: 08800021800014.
Measure the safety distance (SSD) between the X-ray source and the patient. If it's less than 18 cm, the unit is affected.
The DIOX-602 units do not meet the minimum safety distance (SSD) requirement of 18 cm, which could increase radiation exposure risk.
Look for the model number DIOX-602 on the device. The product was sold in the U.S. with the UDI-DI: 08800021800014.
Check the safety distance (SSD) between the X-ray source and the patient. If it's less than 18 cm, contact Digimed for guidance.
We’ve drafted a message you can send Digimed to request your refund or repair — edit it as you like.
Subject: Recall Z-2117-2025 — Digimed Digimed Hello, I own a Digimed that is covered by recall Z-2117-2025 from Digimed. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.