GE Medical Systems recalls certain SIGNA PET/MR MRI scanners that may not have had caster locks checked during maintenance, potentially causing patient table movement during use.
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
GE Medical Systems is recalling specific SIGNA PET/MR MRI scanners that may not have had their patient table caster locks checked during maintenance. This could lead to unintended table movement during scans, risking patient safety.
The patient table's caster locks may not have been properly checked during maintenance, which could allow the table to move unexpectedly during scans. This movement might cause injury to patients or staff during imaging procedures.
Patients and staff using GE SIGNA PET/MR MRI scanners that were maintained by GE HealthCare since March 2023. Those with the specific system IDs listed in the recall notice are most at risk.
Check if your scanner has a system ID number that GE HealthCare performed maintenance on since March 1, 2023. Look for the GTIN numbers listed: 00195278648877, 00840682105378, 00840682123440, 00840682135283.
Verify if your scanner has a system ID that GE HealthCare performed maintenance on since March 1, 2023.
GE HealthCare has become aware that planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed for certain SIGNA PET/MR scanners.
Check if your scanner has a system ID that GE HealthCare performed maintenance on since March 1, 2023. If so, contact GE HealthCare for further instructions.
Yes, the recall includes SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, and other specific models listed in the notice, with some models only available in China.
We’ve drafted a message you can send GE Medical Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-2309-2025 — GE Medical Systems GE Medical Systems Hello, I own a GE Medical Systems that is covered by recall Z-2309-2025 from GE Medical Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.