GE recalls Signa Voyager MRI systems where table caster locks may not have been checked during maintenance since March 2023. Owners should contact GE for a free inspection.
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
GE is recalling Signa Voyager MRI systems that may have missing caster lock checks after maintenance. This could lead to table instability during scans, risking patient or staff injury.
The MRI table's caster locks may not be properly secured after maintenance, causing the table to shift unexpectedly during scans. This could lead to falls or injuries for patients and staff.
MRI facility staff and patients using GE Signa Voyager MRI systems with system IDs checked by GE since March 2023. Those who have had recent maintenance on the system are most at risk.
Check if your system has a GE Signa Voyager model with a system ID that GE HealthCare performed maintenance on since March 1, 2023. GTIN numbers include 00840682108607, 00195278124609, 00195278248923, 00840682118118.
Reach out to GE HealthCare directly to schedule a free inspection of your system's caster locks.
GE HealthCare found that planned maintenance steps to check the MRI table's caster locks may not have been completed for certain systems since March 2023.
Check if your GE Signa Voyager system has a system ID that GE HealthCare performed maintenance on since March 1, 2023. GTIN numbers include 00840682108607, 00195278124609, 00195278248923, 00840682118118.
No, but you should contact GE for a free inspection to ensure the caster locks are secure.
We’ve drafted a message you can send GE Medical Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-2300-2025 — GE Medical Systems GE Medical Systems Hello, I own a GE Medical Systems that is covered by recall Z-2300-2025 from GE Medical Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.