GE Medical Systems Israel recalls Millennium MPR dual-head nuclear medicine gamma cameras due to potential detector fall risk if improperly relocated. Affected units may require repair or replacement to prevent serious injury.
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
GE Medical Systems Israel is recalling certain Millennium MPR gamma cameras used for cardiac imaging. These cameras could have been damaged during relocation, potentially causing a detector to fall and cause serious injury if not properly secured.
The cameras may have been moved without proper detector support, causing stress on the mounting mechanisms. This could compromise the detector's integrity and lead to it falling, potentially causing serious injury.
Healthcare facilities using the Millennium MPR gamma cameras for cardiac imaging are most at risk. Patients and staff could be injured if the camera's detector falls during improper handling or relocation.
The Millennium MPR gamma cameras are used for cardiac imaging. Check for the serial number on the camera or documentation to confirm if it's affected.
Identify the serial number on the camera or its documentation to determine if it's affected by this recall.
GE Medical Systems Israel has become aware that certain Millennium MPR gamma cameras could have been damaged during relocation without adequate detector support, potentially causing a detector to fall and lead to serious injury.
The recall record does not specify a remedy, so contact GE Medical Systems Israel directly for guidance.
Yes, the recall states that improper relocation could cause the detector to fall, potentially leading to serious injury.
We’ve drafted a message you can send GE Medical Systems Israel to request your refund or repair — edit it as you like.
Subject: Recall Z-0299-2026 — GE Medical Systems Israel GE Medical Systems Israel Hello, I own a GE Medical Systems Israel that is covered by recall Z-0299-2026 from GE Medical Systems Israel. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.