ETAC A/S recalls Molift Mover 300 models M1501-M15050 due to bolt deformation causing joint failure and limited mobility. Affected users should check serial numbers listed.
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
ETAC A/S is recalling Molift Mover 300 devices with model numbers M1501 through M15050 because the bolt connecting the lifting bar and arm can deform, leading to limited mobility and equipment failure. This defect affects users who rely on the device for mobility assistance.
The bolt connecting the lifting bar and arm can deform during use, causing limited mobility in the joint. This deformation prevents the device from functioning properly, which may lead to equipment failure when attempting to move the user.
People who own or use Molift Mover 300 devices with model numbers M1501 through M15050 are affected. Those who rely on the device for mobility assistance, such as individuals with limited ability to stand or transfer themselves, are most at risk.
Check if your Molift Mover 300 has model numbers M1501, M1502, M1504, M15050, M15151, M15001, or M15002. The serial numbers listed in the recall notice are also specific to these models.
Review the model number on your Molift Mover 300 to see if it matches the listed numbers (M1501, M1502, M1504, M15050, M15151, M15001, or M15002).
Reach out to ETAC A/S for further assistance or to confirm if your device is affected.
The recall is due to a defect where the bolt connecting the lifting bar and arm can deform, causing limited mobility and equipment failure.
Model numbers M1501, M1502, M1504, M15050, M15151, M15001, and M15002 are affected.
Check the model number on your device against the list of affected models (M1501, M1502, M1504, M15050, M15151, M15001, or M15002).
We’ve drafted a message you can send Molift to request your refund or repair — edit it as you like.
Subject: Recall Z-1309-2026 — Molift Molift Hello, I own a Molift that is covered by recall Z-1309-2026 from Molift. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.