ETAC A/S recalls Molift 4-point sling bars (models 1430115-1430117) due to material deformation causing joint failure. Affected users should stop using until repaired.
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
ETAC A/S recalls Molift 4-point sling bars (models 1430115-1430117) because material deformation can cause the lifting bar to break during assembly or load testing. This defect may prevent the device from working properly, risking injury to users who rely on it for mobility.
The lifting bar can deform between the device and the arm during assembly or load testing, causing limited mobility in the joint. This deformation prevents the device from functioning correctly, potentially leading to falls or injuries during transfers.
Individuals using Molift 4-point sling bars for mobility assistance, particularly those with reduced mobility who depend on the device to transfer between beds, chairs, or wheelchairs. Those who have assembled or tested the device recently are at higher risk.
Check for model numbers 1430115, 1430116, or 1430117 on the product. These Molift 4-point sling bars are accessories for mobile hoists used in mobility assistance.
The device may break during assembly or load testing due to material deformation, so stop using it immediately.
The recall is due to material deformation that can cause the lifting bar to break during assembly or load testing, leading to joint failure and improper device function.
Models 1430115, 1430116, and 1430117.
Look for model numbers 1430115, 1430116, or 1430117 on the Molift 4-point sling bars, which are accessories for mobile hoists used in mobility assistance.
We’ve drafted a message you can send Molift to request your refund or repair — edit it as you like.
Subject: Recall Z-1311-2026 — Molift Molift Hello, I own a Molift that is covered by recall Z-1311-2026 from Molift. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.