Medtronic recalls specific versions of their Vanta Clinician Programmer software for Samsung tablets due to potential delays in surgical procedures. Affected users should contact Medtronic for details.
Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.
Medtronic is recalling two versions of their Vanta Clinician Programmer software for Samsung tablets. The software may cause surgical procedures to run longer than intended, which could impact patient care.
The software may not function correctly during use, potentially leading to extended surgical procedures. This could delay critical medical interventions and affect patient outcomes.
Healthcare professionals using the specific software versions on Samsung tablets. Patients with implanted Vanta neurostimulators are not affected by this recall.
Check for software versions 2.0.2465 or 2.0.2683 on Samsung tablets (model CFN CT900). The product number is A71200. Implanted Vanta neurostimulators are not affected by this recall.
Verify your software version on the Samsung tablet by checking the device settings or the software interface.
No, implanted Vanta Neurostimulator devices are not affected by this recall.
Contact Medtronic directly for instructions on how to update or replace the software.
Check for software versions 2.0.2465 or 2.0.2683 on Samsung tablets (model CFN CT900).
We’ve drafted a message you can send Medtronic to request your refund or repair — edit it as you like.
Subject: Recall Z-1043-2026 — Medtronic Medtronic Hello, I own a Medtronic that is covered by recall Z-1043-2026 from Medtronic. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.