Medline recalls convenience kits with Tego Connectors that may have silicone seal problems, risking blood product leakage or therapy interruption during dialysis.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Medline is recalling convenience kits containing Tego Connectors that may have silicone seals that domes or tear. This could prevent proper blood product injection or withdrawal, causing therapy delays, leaks, or exposure to contaminants.
The silicone seals in the Tego Connectors may dome or tear, blocking the fluid path. This can prevent the user from injecting or withdrawing blood with a syringe, leading to therapy interruptions, leaks, or exposure to contaminants.
Dialysis patients using Medline convenience kits with Tego Connect: those who have the specific kit model and lot numbers listed in the recall. Patients using these kits for blood product handling are most at risk.
Check for the kit model 'Medline DIALYSIS DRSNG CHNG SKU DYNDC3167' with lot numbers 24IME953 or 25CMA708. These kits were sold with UDI-DI (ea) 10195327015954 or UDI-DI (case) 40195327015955.
Identify the kit model and lot numbers from the product label to confirm if it's affected.
Reach out to Medline directly for assistance with the recall process.
The silicone seals in the Tego Connectors may dome or tear, blocking the fluid path and preventing proper blood product injection or withdrawal.
Damaged seals may cause therapy interruptions, fluid leaks, exposure to contaminants, or air infusion into the body.
Check for the kit model 'Medline DIALYSIS DRSNG CHNG SKU DYNDC3167' with lot numbers 24IME953 or 25CMA708.
We’ve drafted a message you can send Medline to request your refund or repair — edit it as you like.
Subject: Recall Z-1468-2026 — Medline Medline Hello, I own a Medline that is covered by recall Z-1468-2026 from Medline. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.