Medline recalls dialysis kits with Tego Connectors due to silicone seal problems that could cause blood leaks or therapy interruptions. Affected kits include specific SKUs and production dates.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Medline is recalling dialysis kits that contain Tego Connectors because the silicone seals might tear or dome. This could lead to blood leaks, interrupted therapy, or exposure to contaminants. Users should check their kit's SKU and production date to see if it's affected.
The silicone seals in Tego Connectors can tear or dome, which may block the fluid path. This prevents proper blood injection or withdrawal with a syringe and could cause leaks, therapy interruptions, or contamination.
Dialysis patients and healthcare workers using Medline DIALYSIS ON/OFF KITS with SKU EBSI1746 are most at risk. Those with kits produced between May 2024 and July 2025 are likely affected.
Check for the SKU EBSI1746 on the kit label. Affected kits were produced with specific dates: 2024051380, 2024052080, 2024052180, 2024052280, 2024071580, 2025061790, or 2025072990.
Look for the SKU EBSI1746 on the kit label and verify the production date against the listed codes.
No, but you should check if your kit is affected before using it. The recall requires verification of the specific SKU and production date.
Damaged seals may cause blood leaks, therapy interruptions, or exposure to contaminants, but the recall does not specify a fix for this issue.
Check the SKU EBSI1746 and production date codes listed: 2024051380, 2024052080, 2024052180, 2024052280, 2024071580, 2025061790, or 2025072990.
We’ve drafted a message you can send Medline to request your refund or repair — edit it as you like.
Subject: Recall Z-1464-2026 — Medline Medline Hello, I own a Medline that is covered by recall Z-1464-2026 from Medline. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.