Pride Mobility Products Corp. recalls Quantum 1450 wheelchairs with software version 2.0.11 due to potential unexpected stops or crashes when using the 'Remind Me Later' feature.
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Pride Mobility Products Corp. is recalling Quantum 1450 wheelchairs with software version 2.0.11. This version may cause the wheelchair to unexpectedly stop while driving or crash when using the 'Remind Me Later' reminder feature, potentially leading to falls or injuries.
If the 'Remind Me Later' option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. This can cause the wheelchair to unexpectedly stop while driving, potentially leading to falls or injuries. If the wheelchair is not being driven, the control system may crash and remain in pre-operational mode.
Users of Quantum 1450 wheelchairs with software version 2.0.11 are affected. People who use the wheelchair while driving or operating it with the 'Remind Me Later' feature are at higher risk.
Check for Quantum 1450 wheelchairs with software version 2.0.11. The product was sold with the software update as a new version.
Install software version 2.0.13 to fix the issue.
Software version 2.0.11 needs to be updated to 2.0.13.
The wheelchair may perform an unexpected park brake stop while driving, which could result in a fall or injury.
Yes, if the wheelchair is not being driven, the control system may crash and remain in pre-operational mode.
We’ve drafted a message you can send Quantum Rehab to request your refund or repair — edit it as you like.
Subject: Recall Z-2847-2026 — Quantum Rehab Quantum Rehab Hello, I own a Quantum Rehab that is covered by recall Z-2847-2026 from Quantum Rehab. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.