Every recall we track for Sysmex America, Inc. — 7 in all — plus the complaints and injury reports the public filed about them.
Sysmex recalls TS-10/TS-10H tube sorters with specific serial numbers due to risk of unintended lift drop. Affected units run software versions 00-17. Contact Sysmex for remedy.
Sysmex recalls PS-10 Sample Preparation System with software versions 1.5 and 1.6 due to false test results from occasional probe wash step omissions during antibody pipetting.
Sysmex recalls PS-10 Sample Preparation System with catalog number BQ716341 due to insufficient antibody causing errors without alarms. Affected units have specific serial numbers listed.
Sysmex recalls CF-70 blood analysis instruments with software mismatch between SP-50 and CF-70 models, which may cause analysis to continue without halting during error conditions.
Sysmex recalls CV-11 Sample Units with serial numbers 11201-15620 due to urine spills causing potential smoke/sparks and fire risk. Affected units may need replacement.
Sysmex recalls XN-Series Fluorocell PLT reagents that may show falsely low platelet counts, affecting blood test accuracy. Affected lots have specific expiration dates.