BioMerieux recalls mini VIDAS systems with Quality Control VIDAS (QCV) due to pump clogging and label displacement causing inaccurate test results. Affected lab equipment may need repair or replacement.
Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS¿ and mini VIDAS¿ system due to a decrease of pipetted volume.
BioMerieux is recalling certain mini VIDAS laboratory testing systems that may cause inaccurate test results due to pump clogging and label displacement. This could lead to incorrect diagnostic outcomes if the equipment is used without proper maintenance.
The pump in the system can clog, and the color-coded SPR label may shift, which reduces the volume of liquid pipetted during tests. This can cause inaccurate test results because the system may not deliver the correct amount of sample for analysis.
Trained laboratory professionals using the mini VIDAS system with Quality Control VIDAS (QCV) for in vitro diagnostics. The risk is primarily for users in veterinary and industrial settings where precise testing is critical.
Check for the product names listed: MINI VIDAS ANALYSER EU, MINI VIDAS ANALYSER US, MINI VIDAS 100-120 VOLT VDE/CE, MINI VIDAS 200-240 VOLT VDE/CE, MINI VIDAS BLUE 110 / 220 V, MINI VIDAS ARKRAY ANALYZER, MINI VIDAS, or MINIVIDAS INDUSTRY US. These models were sold with the Quality Control VIDAS (QCV) for laboratory use.
Identify if your system matches the listed model numbers and product names for the mini VIDAS system with Quality Control VIDAS (QCV).
The recall is due to pump clogging and displacement of the color-coded SPR label, which can cause inaccurate test results by reducing pipetted volume.
The official recall notice does not specify a remedy, so you should check with BioMerieux directly for guidance on next steps.
The product may still function but could produce inaccurate results if the pump clogs or the label shifts, so regular checks are recommended.
We’ve drafted a message you can send BioMerieux to request your refund or repair — edit it as you like.
Subject: Recall Z-2428-2018 — BioMerieux BioMerieux Hello, I own a BioMerieux that is covered by recall Z-2428-2018 from BioMerieux. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.