Allen Medical Systems recalls F-T3SYSTEM T3 surgical table systems with serial numbers A166677-A395123 due to rail clamps that may fail during use, risking patient injury.
The recalling firm has become aware that the rail section clamps used to attach the leg support of the device may fail during use, which could result in critical harm to the patient.
Allen Medical Systems is recalling F-T3SYSTEM T3 surgical table systems because the rail clamps that attach leg supports could fail during use, potentially causing serious harm to patients. This recall affects specific models sold for spinal surgery procedures.
The rail clamps that secure the leg supports can fail while in use, which might cause the table to shift or collapse. This could lead to serious injury or death for patients who are positioned on the table during spinal surgery.
Surgical teams using the F-T3SYSTEM T3 table systems for prone positioning during spinal surgery are most at risk. Patients receiving this equipment during spinal procedures may be affected.
Check for serial numbers starting with A166677 through A395123 on the F-T3SYSTEM T3 surgical table systems. These systems are used for prone positioning during spinal surgery procedures.
Immediately stop using the F-T3SYSTEM T3 surgical table system if it has a serial number from the list provided in the recall.
The F-T3SYSTEM T3 (Trumpf Table Top) System is being recalled for specific serial numbers.
The rail clamps used to attach the leg supports may fail during use, which could result in critical harm to the patient.
Immediately stop using it now and contact Allen Medical Systems for further instructions.
We’ve drafted a message you can send Allen Medical Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-1262-2018 — Allen Medical Systems Allen Medical Systems Hello, I own a Allen Medical Systems that is covered by recall Z-1262-2018 from Allen Medical Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.