MedComp recalls catheter repair kit RMS52001 with UDI 884908039658 due to potential connection failures causing blood loss or air emboli. Affected units include specific lot numbers.
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
MedComp is recalling its catheter repair kit (RMS52001) because it may cause blood loss or air emboli if connections fail. This is serious for patients using dialysis equipment.
The repair kit's female two-part luer connector can fail when attached to hemodialysis tubing, leading to blood loss or air emboli. This is a serious risk during dialysis treatments.
Patients using dialysis equipment with this repair kit are most at risk. Healthcare providers who use the kit for catheter repairs may also be affected.
Check for the product name 'Catheter Repair Kit' with reference number RMS52001 and unique device identifier 884908039658. Affected units have specific lot numbers listed in the recall notice.
Immediately stop using the catheter repair kit if you have it, as it may cause blood loss or air emboli.
The product is the Catheter Repair Kit with reference number RMS52001 and unique device identifier 884908039658.
Affected lot numbers include MCBX400, MCFA150, MCFT460, MCCHOlO, MCDN490, MCCV190, MCCZ680, MCDDlOO, MCDV170, MCDW310.
The repair kit may cause connection failures with hemodialysis tubing, leading to blood loss or air emboli.
We’ve drafted a message you can send MedComp to request your refund or repair — edit it as you like.
Subject: Recall Z-1126-2018 — MedComp MedComp Hello, I own a MedComp that is covered by recall Z-1126-2018 from MedComp. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.