Cook recalls vacuum pumps K-MAR-5200-US model for potential electric shock or burn due to faulty wiring. Affected units may fail to operate safely.
Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
Cook is recalling vacuum pumps (K-MAR-5200-US) that could fail to operate or cause electric shock/burns because of poor wiring. This poses a risk of injury during medical use.
The pump uses crimp connectors that don't meet safety standards. This can cause the device to stop working or lead to electric shock and burns if used improperly.
Healthcare professionals using the vacuum pump for medical procedures like oocyte aspiration are most at risk. Any user of this medical device could be affected.
Check for the model K-MAR-5200-US with serial number GPN G51607. These pumps are used in medical settings for body fluid aspiration.
Ensure the pump operates correctly before use; if it fails, stop using it immediately.
The pump may fail to operate or cause electric shock/burns due to faulty wiring. Do not use until confirmed safe.
Look for the model K-MAR-5200-US and serial number GPN G51607 on the pump.
Check if it operates correctly; if it fails, stop using it immediately. The recall does not specify a return or replacement process.
We’ve drafted a message you can send Cook to request your refund or repair — edit it as you like.
Subject: Recall Z-1460-2018 — Cook Cook Hello, I own a Cook that is covered by recall Z-1460-2018 from Cook. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.