Cenorin recalls HLD Systems 600 Series Washer/Pasteurizers with terminal block overheating risk. Affected models include 610HT (115V/230V) with specific serial numbers. Stop using immediately to prevent fire or burn injuries.
There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.
Cenorin is recalling HLD Systems 600 Series Washer/Pasteurizers due to a supplier quality issue with terminal blocks that may overheat, char, and burn. This poses a fire or burn hazard if the device is used as intended.
The terminal blocks in the device may overheat due to a supplier quality issue, causing them to char and burn. This can lead to fire or burn injuries if the device is used for cleaning and thermally disinfecting medical devices.
Medical device users who own the recalled HLD Systems 600 Series Washer/Pasteurizers with specific serial numbers are most at risk. Healthcare facilities and medical device users should check if their equipment is affected.
Check for the model numbers 610HT (115V or 230V) and specific serial numbers listed in the recall notice. The product is used for cleaning and thermally disinfecting medical devices via full immersion pasteurization.
Verify your device's serial number against the list: 611503 through 611610.
Discontinue use of the device if it has the affected serial numbers to prevent overheating and potential fire or burns.
The recall is due to a supplier quality issue with terminal blocks that may cause overheating, charring, and burning.
The affected models are HLD Systems 600 Series Washer/Pasteurizers with model 610HT (115V or 230V) and specific serial numbers: 611503 through 611610.
Check your device's serial number against the list: 611503, 611504, 611505, 611506, 611601, 611602, 611603, 611604, 611605, 611606, 611607, 611608, 611609, and 611610.
We’ve drafted a message you can send Cenorin to request your refund or repair — edit it as you like.
Subject: Recall Z-2288-2018 — Cenorin Cenorin Hello, I own a Cenorin that is covered by recall Z-2288-2018 from Cenorin. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.