Olympus recalls JF-160F duodenoscopes due to outdated cleaning procedures that could lead to infection if not followed properly.
Issuance of validated, new reprocessing procedures. The new cleaning procedure requires additional recess flushing and forceps elevator raising/lowering steps during precleaning, manual cleaning and manual disinfection.
Olympus is recalling all JF-160F duodenoscopes because their current cleaning process doesn't follow updated safety steps that prevent infections. This matters because improper cleaning could lead to infections if the equipment isn't cleaned correctly.
The current cleaning process for these duodenoscopes doesn't include the new steps required by updated safety guidelines. If staff skip these steps, infections could occur during medical procedures.
Healthcare professionals who use Olympus JF-160F duodenoscopes are affected. Those most at risk are staff handling the equipment without updated cleaning procedures.
Check for the model number JF-160F on the duodenoscope. All units produced by Olympus Corporation of the Americas are included in this recall.
Follow the new reprocessing procedures from Olympus to ensure proper cleaning and disinfection steps.
No, you can continue using it but must follow the updated cleaning procedures to prevent infection risks.
The model number is JF-160F.
The recall does not specify a deadline for updating cleaning procedures.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-2270-2018 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-2270-2018 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.