GE Healthcare recalls CT Goldseal Optima CT 600 systems with improper wiring to UPS causing potential system failure. Affected units have specific serial numbers and system IDs.
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
GE Healthcare is recalling CT Goldseal Optima CT 600 systems that may have improper wiring to the UPS power supply. This could lead to system failure during power outages. Affected units have specific serial numbers and system IDs.
The CT systems may not be properly connected to the Uninterruptible Power Supply (UPS). This could cause the system to fail during power outages, potentially disrupting medical imaging procedures.
Medical facilities that installed GE CT Goldseal Optima CT 600 systems with the specified A! electrical panels. Staff using these systems during power outages are at risk.
Check for serial number 51941YC3 and system ID number 406434660 on the affected CT Goldseal Optima CT 600 systems.
Verify the serial number and system ID on your CT Goldseal Optima CT 600 to see if it matches the affected units.
The CT Goldseal Optima CT 600 with serial number 51941YC3 and system ID number 406434660.
The CT systems may not be properly wired to the partial system UPS, which could cause system failure during power outages.
Check the serial number and system ID to see if it matches the affected units. If it does, contact GE Healthcare for further instructions.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-2583-2018 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-2583-2018 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.