Maquet recalls specific VOLISTA StandOP Surgical Lights (serial numbers from March 2014 to November 2016) due to potential light head detachment during use, risking serious injury to patients or medical staff.
Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.
Maquet is recalling certain VOLISTA StandOP Surgical Lights that could detach from the arm during use, potentially causing serious injury to patients or medical staff. The issue stems from improper assembly of the bracket connecting the light head to the axle fork.
The light head bracket could break due to improper assembly, causing the light head to detach from the arm and hang only by the cable. If the detached light head drops during surgery or procedures, it could impact patients or medical staff.
Medical professionals and patients using the recalled surgical lights are at risk. The hazard is most concerning for those in operating rooms or clinical settings where the lights are used during procedures.
Check for serial numbers starting with ARD568811901 to ARD568821961, manufactured between March 10, 2014 and November 17, 2016. The lights are used in surgical suites, clinics, and doctor's offices.
Look for serial numbers starting with ARD568811901 to ARD568821961 on the product.
If the serial number matches the recalled range, stop using the light immediately to prevent detachment risks.
The light head bracket could break, causing the light head to detach from the arm and hang only by the cable. If it drops during use, it could injure patients or medical staff.
Check the serial number to see if it's in the recalled range. If it is, stop using it immediately and contact Maquet for further instructions.
These lights were manufactured from March 10, 2014, to November 17, 2016.
We’ve drafted a message you can send Maquet to request your refund or repair — edit it as you like.
Subject: Recall Z-2502-2018 — Maquet Maquet Hello, I own a Maquet that is covered by recall Z-2502-2018 from Maquet. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.