Degania recalls nonsterile thermistor Foley catheters that may show lower body temperature than actual patient temperature during surgery or post-op care. This could lead to inaccurate temperature monitoring.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Degania is recalling nonsterile thermistor Foley catheters that may show lower body temperature readings than actual patient temperature. This could cause inaccurate temperature monitoring during surgery or after surgery, potentially leading to medical errors.
The catheters contain defective temperature sensors that may show lower body temperature readings than the actual patient temperature. This could lead to inaccurate temperature monitoring during surgery or after surgery, potentially causing medical errors if the temperature is misinterpreted.
Patients undergoing surgery or post-surgical care who use these catheters for temperature monitoring are affected. Healthcare providers using these catheters for temperature monitoring during medical procedures are at risk.
Check for the catalogue number 102201101663MD on the catheter packaging. These catheters were sold with specific serial numbers: V17012299, V17012518, V17012896, V17013267, V17013268, V17013272, V17023765, V17025022, V17025023, V17026861.
Look for the catalogue number 102201101663MD on the packaging or label.
The recall states these catheters may show lower body temperature readings than actual patient temperature, so they are not safe for use without checking the recall.
The recall does not specify a remedy, so contact the manufacturer for guidance.
Check for the catalogue number 102201101663MD and the specific serial numbers listed: V17012299, V17012518, V17012896, V17013267, V17013268, V17013272, V17023765, V17025022, V17025023, V17026861.
We’ve drafted a message you can send Degania to request your refund or repair — edit it as you like.
Subject: Recall Z-2778-2018 — Degania Degania Hello, I own a Degania that is covered by recall Z-2778-2018 from Degania. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.