Zimmer Biomet recalls Affixus Hip Fracture Nail Right 125mm model due to set screw not advancing or reversing properly. Affected units may require replacement.
The set screw is not able to advance or reverse from the original position in the nail.
Zimmer Biomet is recalling a specific hip fracture nail model because its set screw doesn't work correctly. This could prevent proper fixation of the hip fracture, potentially causing complications after surgery.
The set screw in the nail cannot move forward or backward from its starting position. This may prevent the nail from securely fixing the broken hip bone, which could lead to complications during healing.
Surgical patients who received this specific hip fracture nail model during surgery. Patients using the product for medical purposes are most at risk.
Check for the model name 'Affixus Hip Fracture Nail Right 125mm' and item number 814311340. The product was sold in specific quantities across different regions.
Identify the model name and item number to confirm if your product is affected.
The recall states the set screw may not advance or reverse properly, which could affect surgical fixation. Affected units should not be used.
Look for the model name 'Affixus Hip Fracture Nail Right 125mm' and item number 814311340.
The recall does not specify a remedy, so contact Zimmer Biomet directly for guidance.
We’ve drafted a message you can send Zimmer Biomet to request your refund or repair — edit it as you like.
Subject: Recall Z-3059-2018 — Zimmer Biomet Zimmer Biomet Hello, I own a Zimmer Biomet that is covered by recall Z-3059-2018 from Zimmer Biomet. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.