Zimmer Biomet recalls Affixus Hip Fracture Nail Left 130 (13mm x 460mm) due to set screw not advancing or reversing properly. Affected units may need repair or replacement.
The set screw is not able to advance or reverse from the original position in the nail.
Zimmer Biomet is recalling Affix: Hip Fracture Nail Left 130 models because the set screw cannot move properly. This could lead to improper fixation of the hip fracture, risking further injury during surgery or recovery.
The set screw in the nail fails to advance or reverse from its original position. This could cause the nail to become improperly fixed in the bone, potentially leading to displacement or failure during healing.
Surgical patients who received the Affixus Hip Fracture Nail Left 130 (13mm x 460mm) model during surgery. Patients using this specific nail for hip fracture repair are most at risk.
Check for the product name 'Affixus Hip Fracture Nail Left 130' with dimensions 13 mm x 460 mm. The item number 814613460 is also associated with this model.
Reach out to Zimmer Biomet directly for guidance on next steps for affected units.
The recall states the set screw may not advance or reverse properly, which could lead to improper fixation. However, the hazard is not described as causing immediate injury in normal use.
Check for the product name 'Affixus Hip Fracture Nail Left 130' with dimensions 13 mm x 460 mm and item number 814613460.
The recall does not specify a remedy, so contact Zimmer Biomet directly for guidance.
We’ve drafted a message you can send Zimmer Biomet to request your refund or repair — edit it as you like.
Subject: Recall Z-3173-2018 — Zimmer Biomet Zimmer Biomet Hello, I own a Zimmer Biomet that is covered by recall Z-3173-2018 from Zimmer Biomet. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.