Cambridge Sensors recalls specific Microdot Xtra blood glucose test strips due to improper fit causing delayed test results. Affected lots listed in the notice.
The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.
Cambridge Sensors is recalling certain Microdot Xtra blood glucose test strips because they don't fit properly in the blood glucose meter, causing delays in test results. This could lead to inaccurate readings if users rely on these strips for monitoring.
The faulty test strips do not fit correctly into the blood glucose meter's inlet port due to a manufacturing issue with the cutting machine. This causes delays in getting test results, which could impact accurate blood glucose monitoring.
People who own or use the specific Microdot Xtra blood glucose test strips with the listed lot numbers are affected. Those who rely on these strips for blood glucose monitoring are most at risk of delayed results.
Check the lot number on the test strip packaging. Affected lots include 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702.
Look for the lot number on the test strip packaging to determine if it is affected.
The faulty cutting machine did not correctly notch the test strips, causing them to not fit into the blood glucose meter's inlet port and resulting in delayed test results.
Check the lot number on the test strip packaging. If it matches the listed affected lots, stop using those strips and contact Cambridge Sensors for further instructions.
No, the recall is only for the test strips. The blood glucose meter itself is not affected.
We’ve drafted a message you can send Cambridge Sensors to request your refund or repair — edit it as you like.
Subject: Recall Z-2357-2019 — Cambridge Sensors Cambridge Sensors Hello, I own a Cambridge Sensors that is covered by recall Z-2357-2019 from Cambridge Sensors. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.