Oridion Medical recalls Capnostream20 N W/PRINTER models with resettable alarm settings. Users may need to adjust settings after power-off to avoid incorrect patient monitoring.
The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.
Oridion Medical is recalling certain Capn:stream20 N W/PRINTER models because their alarm and nurse call settings can reset to factory defaults when the monitor is powered off. This could lead to incorrect patient monitoring if settings aren't manually adjusted after power-off.
When the monitor is powered off, the alarm and nurse call settings may reset to factory defaults. This could cause incorrect patient monitoring if the settings aren't manually adjusted after power-off. The risk is low because the device is designed for medical use and settings reset only during power-off cycles.
Healthcare professionals using the Capnostream20 N W/PRINTER models listed in the recall. Patients under continuous monitoring with these devices are at risk if settings reset unexpectedly.
Check for the model numbers B400002649, B400002619, B400002624, B400002632, or B400002641 on the device. These models were sold with the Capnostream20 N W/PRINTER CYVITAL.
After powering off the monitor, verify that alarm and nurse call settings have been manually adjusted to avoid resetting to factory defaults.
Yes, you should check the settings after powering off the monitor to ensure they haven't reset to factory defaults.
The affected models are B400002649, B400002619, B400002624, B400002632, and B400002641.
The recall does not provide a specific fix; users must manually adjust settings after power-off to prevent resets.
We’ve drafted a message you can send Oridion Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0049-2019 — Oridion Medical Oridion Medical Hello, I own a Oridion Medical that is covered by recall Z-0049-2019 from Oridion Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.