Olympus recalls BF-MP60 bronchovideoscopes with non-Olympus accessories that may loosen the instrument channel port, risking injury. Affected units include all lot codes.
The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Olympus is recalling all Bronchovideoscope BF-MP60 models because attaching non-Olympus accessories can cause the instrument channel port to loosen. This could lead to injury if the port becomes loose during use.
Non-Olympus accessories attached to the instrument channel port may apply more force than expected, causing the port to loosen. This could lead to injury if the port becomes loose during medical procedures.
Healthcare professionals using the Bronchovideoscope BF-MP60 are most at risk. Patients and staff could be affected if the port loosens during procedures.
Check for the Bronchovideoscope BF-MP60 model with any lot code. The product was sold by Olympus Corporation of the Americas.
Remove any non-Olympus accessories attached to the instrument channel port.
Confirm the lot code on the product to determine if it is affected.
The recall only affects non-Olympus accessories. Olympus accessories should be safe.
Check for non-Olympus accessories and verify the lot code to see if it is affected.
The recall record does not mention a replacement or remedy.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-3295-2018 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-3295-2018 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.