Olympus recalls BF-XT40 bronchoscopes with non-Olympus accessories that may loosen the instrument channel port, risking injury during use.
The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Olympus is recalling all Bronchofiberscope BF-XT40 models due to a risk that non-Olympus accessories could loosen the instrument channel port, potentially causing injury during medical procedures.
Non-Olympus accessories attached to the instrument channel port can apply more force than expected, causing the port to loosen. This could lead to injury during medical procedures if the port becomes dislodged.
Medical professionals using the Bronchofiberscope BF-XT40 are most at risk. Patients undergoing procedures with this device may also be affected if non-Olympus accessories are used.
Check for the Bronchofiberscope BF-XT40 model with any lot code. The recall applies to all units sold by Olympus Corporation of the Americas.
Ensure all accessories attached to the instrument channel port are Olympus-branded to avoid loosening risks.
Yes, Olympus recalls all Bronchofiberscope BF-XT40 models with any lot code.
The port loosening could cause injury during medical procedures if the instrument channel becomes dislodged.
No, but you should check for non-Olympus accessories and ensure they are Olympus-branded to prevent port loosening.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-3309-2018 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-3309-2018 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.