GE Healthcare recalls Dash 3000 Patient Monitoring Systems that may restart simultaneously during network overload, potentially disrupting patient monitoring. Affected units were sold to hospitals and clinics.
The patient monitors may simultaneously restart as designed if all are connected to the same network and a network overload occurs for a prolonged time.
GE Healthcare is recalling its Dash 3000 Patient Monitoring Systems because they may restart at the same time if connected to the same network and a network overload occurs for a prolonged period. This could disrupt patient monitoring and pose risks during critical medical situations.
The patient monitors may restart simultaneously when connected to the same network and a network overload occurs for an extended period. This could cause interruptions in patient monitoring, potentially affecting critical health data during medical procedures.
Healthcare facilities using the Dash 3000 Patient Monitoring Systems for adult, pediatric, and neonatal patients are most at risk. Hospitals and clinics that have purchased these monitors are likely affected.
The Dash 3000 Patient Monitoring System is used for monitoring physiologic parameters in adult, pediatric, and neonatal patients. Affected units were sold to hospitals and clinics.
Ensure all monitors are not connected to the same network if they are part of a system that may experience prolonged network overloads.
The recall applies to hospitals and clinics that have purchased the Dash 3000 Patient Monitoring Systems for adult, pediatric, and neonatal patients.
The official recall notice does not specify a remedy or action steps for affected units.
The recall states that the monitors may restart simultaneously during network overload, but does not mention specific injury risks.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0519-2019 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0519-2019 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.