Fort Defiance Industries recalls Automated Field Steam Sterilizers REF P2131 due to potential ground connection failure. Affected units may have unreliable fault current return, risking electrical hazards.
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Fort Defiance Industries is recalling their Automated Field Steam Sterilizers with model REF P2131 because a water recovery system connector might lose ground connection. This could cause electrical issues that might lead to safety risks if the device is used improperly.
The 120V Amphenol connector in the water recovery system might lose ground connection, creating an unreliable path for fault current. This could lead to electrical malfunctions, potentially causing shocks or equipment damage if the device is used without proper grounding.
Users of the Automated Field Steam Sterilizer REF P2131 are affected. Those who use the device in environments where electrical safety is critical, like medical or industrial settings, are at higher risk.
Check for the model REF P2131 on the device. All units with this model number are affected, including those with any serial number or GTIN code 01 00862992000327.
Look for the model number REF P2131 on the device label or documentation.
Reach out to Fort Defiance Industries for specific guidance on next steps.
The recall indicates a potential electrical hazard, so it's not recommended to use the device until the manufacturer provides guidance.
Check for the model number REF P2131 on the device. All units with this model are affected, including those with any serial number or GTIN code 01 00862992000327.
The recall does not specify a remedy, so contact Fort Defiance Industries directly for instructions.
We’ve drafted a message you can send Fort Defiance Industries to request your refund or repair — edit it as you like.
Subject: Recall Z-2983-2026 — Fort Defiance Industries Fort Defiance Industries Hello, I own a Fort Defiance Industries that is covered by recall Z-2983-2026 from Fort Defiance Industries. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.