Medtronic recalls CareLink Encore 29901 programmers due to potential cybersecurity risks allowing unauthorized software updates. All units are affected.
Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
Medtronic is recalling all CareLink Encore 29901 programmers because vulnerabilities could let malicious actors update the device with non-Medtronic software during software downloads. This could compromise device security and patient safety.
The vulnerability allows someone with malicious intent to update the programmer with unauthorized software during a software download process. This could lead to device malfunction or unintended health impacts.
Patients using Medtronic CareLink Encore 29901 programmers for heart rhythm monitoring are affected. Those with medical devices connected to the programmer are at higher risk.
The device is the Medtronic CareLink Encore 29901 Programmer. All serial numbers are affected, so any unit purchased by Medtronic is included.
Identify the serial number on the device to confirm it is affected by this recall.
The recall is for the CareLink Encore 29901 programmer, not the heart device itself. Patients using this programmer for monitoring are affected.
The official recall does not specify security measures. Contact Medtronic directly for guidance.
The recall does not include a stop-use instruction. The remedy is not stated in the record.
We’ve drafted a message you can send Medtronic to request your refund or repair — edit it as you like.
Subject: Recall Z-0407-2019 — Medtronic Medtronic Hello, I own a Medtronic that is covered by recall Z-0407-2019 from Medtronic. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.