Ortho-Clinical Diagnostics recalls VITROS 4600 Chemistry System versions 3.3.1 and below due to software anomaly that may incorrectly extend reagent stability time.
Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
Ortho-Clinical Diagnostics is recalling VITROS 4600 Chemistry System versions 3.3.1 and below. This software issue could cause reagents to stay stable longer than they should, potentially leading to inaccurate test results.
The software anomaly may incorrectly extend the on-analyzer stability time for loaded reagents. This could lead to reagents being used beyond their safe stability period, resulting in inaccurate test results.
Healthcare facilities using the VITROS 4600 Chemistry System with software version 3.3.1 or below are affected. Laboratory staff and patients using these systems are at risk of inaccurate test results.
Check if your VITROS 4600 Chemistry System has software version 3.3.1 or below. Product code 6802445, UDI 10758750012343, and serial numbers J46000108-J46001132 are associated with this recall.
Verify your VITROS 4600 Chemistry System software version to see if it is 3.3.1 or below.
Software version 3.3.1 and below.
Check the product code (6802445), UDI (10758750012343), or serial numbers (J46000108-J46001132) to confirm if your system is affected.
The recall record does not specify a remedy, so no action is required beyond checking your software version.
We’ve drafted a message you can send Ortho-Clinical Diagnostics to request your refund or repair — edit it as you like.
Subject: Recall Z-0653-2019 — Ortho-Clinical Diagnostics Ortho-Clinical Diagnostics Hello, I own a Ortho-Clinical Diagnostics that is covered by recall Z-0653-2019 from Ortho-Clinical Diagnostics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.