Civco recalls needle guide starter kits with lot numbers A000012-A050663, M000880-M995980, and P022503 due to potential contamination from non-sterile brackets.
See the official notice for the full hazard description.
Civco is recalling a needle guide starter kit that uses non-sterile brackets. This poses a risk of contamination during medical procedures, which could lead to infections.
The kit includes a reusable non-sterile bracket that may not be properly sterilized. This can lead to contamination during medical procedures if the bracket is not cleaned or sterilized correctly.
Healthcare professionals using the kit for medical procedures are most at risk. Patients receiving procedures with this kit could be exposed to contamination.
Check the lot numbers on the box: A000012-A050663, M000880-M995980, or P022503. The kit includes brackets with depths of 1.5cm, 2.5cm, 3.5cm, and 4.5cm.
Look for lot numbers on the box: A000012-A050663, M000880-M995980, or P022503.
If the brackets no longer have product labeling, they should be discarded to prevent contamination.
The kit includes non-sterile brackets that may not be properly sterilized, so it is not safe for medical use without proper cleaning.
Lot numbers A000012-A050663, M000880-M995980, and P022503 are affected.
Check the box for lot numbers: A000012-A050663, M000880-M995980, or P022503. Also, brackets without product labeling are recalled.
We’ve drafted a message you can send Civco to request your refund or repair — edit it as you like.
Subject: Recall Z-0571-2019 — Civco Civco Hello, I own a Civco that is covered by recall Z-0571-2019 from Civco. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.