Philips recalls Intera 1.5T MRI systems with model 781295 due to potential ceiling speaker falling and being pulled by magnetic field. Affected units may need repair or replacement.
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Philips is recalling its Intera 1.5T High Performance MRI systems with model number 781295. The risk is that ceiling speakers could fall and be pulled toward the MRI machine by the strong magnetic field, potentially causing injury.
The MRI machine's strong magnetic field can attract metal objects. If ceiling speakers contain metal parts, they might become dislodged and pulled toward the machine, potentially causing injury to nearby people.
Medical facilities using Philips Intera 1.5T MRI systems with model 781295 are most at risk. Patients and staff near the MRI machine during operation could be affected.
Check for the model number 781295 on the Intera 1.5T High Performance MRI system. This recall applies to all systems sold by Philips Medical Systems Nederlands.
Look for the model number 781295 on the MRI system.
Reach out to Philips Medical Systems for guidance on next steps.
Yes, ceiling speakers could fall and be pulled by the magnetic field, potentially causing injury.
Check for the model number 781295 on the Intera 1.5T High Performance MRI system.
Contact Philips Medical Systems for guidance on next steps.
We’ve drafted a message you can send Philips Medical Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-0908-2019 — Philips Medical Systems Philips Medical Systems Hello, I own a Philips Medical Systems that is covered by recall Z-0908-2019 from Philips Medical Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.