Philips recalls Panorama 1.0T MRI systems with model 781250 due to risk of ceiling speakers falling and being pulled by magnetic fields. Affected units must be inspected and corrected.
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Philips is recalling its Panorama 1.0T MRI systems with model number 781250 because ceiling speakers could fall and be pulled toward the MRI machine by magnetic fields. This hazard could cause injury if the speakers fall and strike someone.
The magnetic field of the MRI system can attract metal objects, including ceiling speakers. If the speakers fall, they could strike people in the room, causing injury.
Healthcare facilities using Philips Panorama 1.0T MRI systems with model number 781250 are affected. Staff and patients near the MRI room are at risk if the speakers fall.
Check for the model number 781250 on the Panorama 1.0T MRI system. This recall applies to all systems sold by Philips Medical Systems Nederlands.
Look for the model number 781250 on the Panorama 1.0T MRI system.
The ceiling speakers could fall and be pulled toward the MRI machine by magnetic fields, potentially causing injury.
Check the model number on your system. If it's 781250, contact Philips Medical Systems Nederlands for instructions.
Yes, if the speakers fall and strike someone, it could cause injury.
We’ve drafted a message you can send Philips Medical Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-0898-2019 — Philips Medical Systems Philips Medical Systems Hello, I own a Philips Medical Systems that is covered by recall Z-0898-2019 from Philips Medical Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.