Philips recalls Achieva 1.5T MRI systems with model 781196 due to risk of ceiling speakers falling and being pulled by the magnetic field. Affected units may need inspection or repair.
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Philips is recalling certain Achieva 1.5T MRI systems with model number 781196 because ceiling speakers could fall and be pulled by the magnetic field. This poses a risk of injury if the speakers fall and hit someone.
The magnetic field in MRI systems can attract metal objects. In this case, ceiling speakers might detach and be pulled toward the MRI machine, potentially causing injury if they strike a person.
Healthcare facilities using Philips Achieva 1.5T MRI systems with model number 781196 are most at risk. Patients and staff near the MRI room during operation could be affected if speakers fall.
Check for the model number 781196 on the Achieva 1.5T High Performance MRI system. These systems were sold by Philips Medical Systems Nederlands.
Look for the model number 781196 on the Achieva 1.5T MRI system.
The ceiling speakers could fall and be pulled by the magnetic field, posing a risk of injury.
Check the model number on your system. If it's 781196, contact Philips Medical Systems for guidance.
The recall states a potential risk of injury if the speakers fall and hit someone, but no specific injuries were reported.
We’ve drafted a message you can send Philips Medical Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-0896-2019 — Philips Medical Systems Philips Medical Systems Hello, I own a Philips Medical Systems that is covered by recall Z-0896-2019 from Philips Medical Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.