Stryker recalls high flow tube sets with real-time pressure sensing (RTP) that leak and detach from the cassette. Affected units have specific lot numbers and reference codes. Stop using immediately to prevent potential injury.
The heated tube sets were consistently leaking and detaching from the cassette.
Stryker is recalling high flow tube sets with real-time pressure sensing that leak and detach from the cassette. This could cause the tubes to come loose during use, potentially leading to injury or equipment failure.
The heated tube sets consistently leak and detach from the cassette during use. This can cause the tubes to become dislodged, potentially leading to equipment malfunction or injury if the tubes are not properly secured.
Users of the Stryker PNEUMO SURE High Flow Insufflator who own the recalled tube sets are most at risk. This includes medical professionals and patients using the equipment in clinical settings.
Check for the reference code REF 0620-040-680 on the tube set. Affected units have specific lot numbers listed in the recall notice, including 18G0273 through 56006606.
Discontinue use of the recalled tube set to prevent detachment and potential injury.
The recalled tube set is the Stryker High Flow Tube Set with Real-Time Pressure Sensing (RTP), reference code REF 0620-040-680.
Check the reference code on the tube set and compare it to the listed lot numbers in the recall notice (18G0273 through 56006606).
Stop using the tube set immediately and contact Stryker for further instructions.
We’ve drafted a message you can send Stryker to request your refund or repair — edit it as you like.
Subject: Recall Z-1727-2019 — Stryker Stryker Hello, I own a Stryker that is covered by recall Z-1727-2019 from Stryker. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.