Medtronic Vascular recalls SHERPA NX ACTIVE GUIDING CATHETER models for potential wire exposure. Learn what to do if you own this device.
There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters.
Medtronic Vascular is recalling a specific type of cardiovascular catheter due to a risk of internal wires becoming exposed. This could lead to complications during medical procedures if the catheter is used improperly.
The catheter's primary segment material may degrade, exposing the underlying stainless-steel braid wires. This could cause complications during procedures if the wires become tangled or kinked.
Patients and medical professionals who use the recalled catheter for cardiovascular procedures. Those most at risk are individuals with specific catheter model numbers.
Check for the model name 'SHERPA NX ACTIVE GUIDING CATHETER', size 6F JR4.5, diameter .070", and reference number SA6JR45. The product was sold with a GTIN: 00613994821829.
Identify the model, size, and reference number to confirm if your catheter is recalled.
The recall states there is a potential for wire exposure, but it does not specify that the catheter is unsafe when used correctly.
The recall record does not provide a specific remedy, so contact Medtronic Vascular directly for guidance.
Look for the model name 'SHERPA NX ACTIVE GUIDING CATHETER', size 6F JR4.5, diameter .070", and reference number SA6JR45.
We’ve drafted a message you can send Medtronic Vascular to request your refund or repair — edit it as you like.
Subject: Recall Z-2630-2019 — Medtronic Vascular Medtronic Vascular Hello, I own a Medtronic Vascular that is covered by recall Z-2630-2019 from Medtronic Vascular. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.