Medtronic recalls specific SHERPA NX Active Guiding Catheters for potential wire exposure risk during cardiovascular procedures.
There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters.
Medtronic is recalling certain SHERPA NX Active Guiding Catheters used in cardiovascular procedures. The catheters may expose underlying steel wires, which could cause serious complications if used improperly.
The catheter's primary segment material may break off, exposing the stainless-steel braid wires. This could lead to complications like tissue damage or blockages if the wires become embedded in blood vessels.
Cardiovascular doctors and medical staff who use the recalled catheters during procedures. Patients receiving these catheters are at risk if the catheter malfunctions.
Check for the model SHERPA NX ACTIVE GUIDING CATHETER, 6F ECR3.5 SH, .070", REF SA6ECR35SH. The product was sold with GTIN 00613994824837. All lot and serial numbers are included in the recall.
Verify the model number and serial number against the recall list provided by Medtronic.
The recall indicates a potential risk, so it is not safe to use without checking the recall list.
The recall does not specify a remedy, so contact Medtronic directly for guidance.
Check the model number and serial number against the recall list: SHERPA NX ACTIVE GUIDING CATHETER, 6F ECR3.5 SH, .070", REF SA6ECR35SH.
We’ve drafted a message you can send Medtronic Vascular to request your refund or repair — edit it as you like.
Subject: Recall Z-2714-2019 — Medtronic Vascular Medtronic Vascular Hello, I own a Medtronic Vascular that is covered by recall Z-2714-2019 from Medtronic Vascular. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.