Thermo Fisher Scientific recalls 28L precision baths with model TSGP28 due to potential overheating from incorrect assembly. Affected units have specific serial numbers.
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Thermo Fisher is recalling 28L precision baths with model number TSGP28 that may have been assembled incorrectly. This could cause the bath to overheat beyond the user's set temperature, potentially leading to unsafe conditions.
The bath may have been assembled incorrectly, causing it to malfunction and raise the temperature of the fluid in the reservoir above the user's selected temperature setting. This overheating could lead to unsafe conditions if the bath is used without proper monitoring.
Users of the Thermo Scientific General Purpose Dynamic Bath Precision 28L GP Bath with model number TSG:GP28 are affected. Laboratory technicians and researchers using these baths for scientific work are most at risk.
Check for the model number TSGP28 on the bath. Affected units have specific serial numbers: 300246819, 300247423, and 300247424. The product was sold by Thermo Fisher Scientific.
Look for the model number TSGP28 on the bath to confirm if it's affected.
Contact Thermo Fisher Scientific for assistance with the recall if you have the affected model.
The model number is TSGP28.
Serial numbers 300246819, 300247423, and 300247424 are affected.
Check for the model number TSGP28 and the specific serial numbers listed in the recall notice.
We’ve drafted a message you can send Thermo Fisher to request your refund or repair — edit it as you like.
Subject: Recall Z-1872-2019 — Thermo Fisher Thermo Fisher Hello, I own a Thermo Fisher that is covered by recall Z-1872-2019 from Thermo Fisher. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.