Abbott recalls Alinity c Cuvette Segments (List 04S47-01) due to potential inadequate sample dispensing. Affected units may cause inaccurate test results. Contact Abbott for details.
Individual cuvettes within the Alinity c Cuvette Segment may become seated lower than the designed height. This may result in inadequate dispense into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom.
Abbott is recalling Alinity c Cuvettes that may not dispense samples properly, leading to inaccurate test results. This could affect the reliability of medical testing results.
The cuvettes may sit lower than designed, causing the sample probe to fail contacting the cuvette bottom. This leads to inadequate sample dispensing and potentially inaccurate test results.
Users of the Alinity c system for medical testing are affected. Those who rely on accurate test results from this device are at higher risk.
Check for Alinity c Cuvette Segments with List Number 04S4:7-01. These were sold with the Alinity c system for medical testing.
Identify if your Alinity c Cuvette Segments have List Number 04S47-01.
The cuvettes may sit lower than designed, causing the sample probe to fail contacting the cuvette bottom, leading to inadequate sample dispensing.
The recall record does not specify a remedy or action steps.
Check for Alinity c Cuvette Segments with List Number 04S47-01 sold with the Alinity c system.
We’ve drafted a message you can send Abbott to request your refund or repair — edit it as you like.
Subject: Recall Z-1709-2019 — Abbott Abbott Hello, I own a Abbott that is covered by recall Z-1709-2019 from Abbott. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.