Maquet recalls specific models of small 1/4 tubing sets for extracorporeal circulation due to potential sterility issues during sterilization. Affected units have batch numbers listed in the recall notice.
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Maquet is recalling certain small 1/4 tubing sets for extracorporeal circulation because they may not maintain sterility during sterilization. This could lead to infections if the equipment is used without proper sterility.
During sterilization, the tubing's closed stopcock design may trap humidity and ethylene oxide gas, preventing proper sterility assurance. This could lead to contamination if the equipment is used without adequate sterility.
Healthcare professionals using these tubing sets for extracorporeal circulation systems are most at risk. Patients receiving such treatments could be affected if the equipment is not properly sterilized.
Check for the model number 'Getinge-BEQ-T 9410 Lurie Chldns Small 1/4' and batch numbers 3000060035, 3000065494, 3000067883, 3000071809, 3000074480, 3000075340, or 3000077550.
Look for the model number 'Getinge-BEQ-T 9410 Lurie Chldns Small 1/4' and batch numbers listed in the recall notice.
The tubing set may not maintain sterility during sterilization due to its closed stopcock design, which can trap humidity and ethylene oxide gas.
Batch numbers 3000060035, 3000065494, 3000067883, 3000071809, 3000074480, 3000075340, and 3000077550 are affected.
The recall does not specify a remedy, so there is no immediate action required beyond checking if your product is affected.
We’ve drafted a message you can send Maquet to request your refund or repair — edit it as you like.
Subject: Recall Z-1620-2019 — Maquet Maquet Hello, I own a Maquet that is covered by recall Z-1620-2019 from Maquet. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.