Isopure Corp recalls Finish Thompson Centrifugal Pump Model #110192-3 spare parts due to potential uneven magnet encapsulation causing corrosion. Affected units have specific serial numbers listed.
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
Isopure Corp is recalling Finish Thompson Centrifugal Pump spare parts model #110192-3 because a molding issue may cause uneven magnet wear and corrosion. This could lead to pump failure if the magnet is exposed to solution.
During manufacturing, the magnet ring may not be evenly encapsulated in plastic, causing partial displacement. This leads to uneven wear and potential exposure of the magnet to solution, which can cause corrosion and iron contamination.
Spare part users for Finish Thompson Centrifugal Pumps, particularly those with the listed serial numbers. Most at risk are technicians or maintenance staff handling these pumps.
Check if your Finish Thompson Centrifugal Pump spare part has serial numbers: 178323B18, 178751C18, 178749C18, 180214D18, 180213D18, 180208D18, 178747C18, or 178750C18.
Verify if your spare part has one of the listed serial numbers: 178323B18, 178751C18, 178749C18, 180214D18, 180213D18, 180208D18, 178747C18, or 178750C18.
The pump may have uneven magnet encapsulation during manufacturing, leading to partial displacement, uneven wear, and potential corrosion if the magnet is exposed to solution.
Check if your spare part has one of the listed serial numbers. If it does, contact Isopure Corp for further instructions.
The recall record does not mention a remedy or fix.
We’ve drafted a message you can send Isopure Corp to request your refund or repair — edit it as you like.
Subject: Recall Z-0823-2020 — Isopure Corp Isopure Corp Hello, I own a Isopure Corp that is covered by recall Z-0823-2020 from Isopure Corp. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.