Inovo recalls Single Luman Conserver oxygen regulators with serial numbers starting with C0019802 to C0020216. The pressure gauge may dislodge from the threaded mount, creating an oxygen leak. Users should stop using the device immediately.
The pressure gauge (manometer) may dislodge from the threaded insertion/mount, creating a leak.
Inovo is recalling Single Luman Conserver oxygen regulators that may have a pressure gauge that dislodges from the threaded mount, causing oxygen leaks. This could lead to unsafe oxygen flow during medical use.
The pressure gauge (manometer) in the device can become loose or dislodge from the threaded insertion point. This may create an oxygen leak, which could lead to unsafe oxygen delivery during medical use.
Patients using medical oxygen therapy with the recalled Single Luman Conserver regulators are most at risk. Owners of the specific serial numbers listed in the recall are affected.
Check for serial numbers starting with C0019802 to C0020216, including both the IN6505-L-Blue and ML6505-L models. The device is intended for regulating oxygen flow from an oxygen tank.
Immediately stop using the recalled Single Luman Conserver oxygen regulator if you have the affected serial numbers.
The product is the Inovo Single Luman Conserver, a device that regulates oxygen flow from an oxygen tank.
Serial numbers starting with C0019802 to C0020216, including both IN6505-L-Blue and ML6505-L models.
Check the serial number on your device. If it starts with C0019802 to C0020216 (including IN6505-L-Blue and ML6505-L models), it is affected.
We’ve drafted a message you can send Inovo to request your refund or repair — edit it as you like.
Subject: Recall Z-2099-2019 — Inovo Inovo Hello, I own a Inovo that is covered by recall Z-2099-2019 from Inovo. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.