Medtronic recalls 0-arm TM 1000 Imaging Systems with serial numbers Bl-700-00027-XXX and B1-700-00028-XXX where XXX indicates system voltage. If the charger board is faulty, the system may not take X-rays or move properly, affecting surgery.
The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced
Medtronic is recalling certain 0-arm TM 1000 Imaging Systems that may not take X-rays or move properly if the charger board is faulty. This could disrupt surgery by preventing the system from functioning correctly in the operating room.
The charger board may not function properly, causing the batteries to not receive a full recharge. This prevents the system from taking X-rays, opening the gantry, or moving between locations, which could disrupt surgical procedures.
Healthcare facilities using the recalled 0-arm TM 1000 Imaging Systems with specific serial numbers. Surgeons and medical staff who rely on these systems for imaging during procedures are most at risk.
Check the serial numbers listed in the recall notice. The affected models have serial numbers starting with Bl-700-00027-XXX or B1-700-00028-XXX where XXX indicates the system voltage (e.g., 100V, 120V, 230V).
Look for serial numbers starting with Bl-700-00027-XXX or B1-700-00028-XXX where XXX indicates the system voltage (e.g., 100V, 120V, 230V).
Medtronic 0-arm TM 1000 Imaging Systems
Serial numbers starting with Bl-700-00027-XXX or B1-700-00028-XXX where XXX indicates the system voltage (e.g., 100V, 120V, 230V).
The charger board may not function properly, causing the batteries to not receive a full recharge and preventing the system from taking X-rays or moving properly.
We’ve drafted a message you can send Medtronic to request your refund or repair — edit it as you like.
Subject: Recall Z-1932-2019 — Medtronic Medtronic Hello, I own a Medtronic that is covered by recall Z-1932-2019 from Medtronic. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.