Maquet recalls VOLISTA StandOP Surgical Lights with potential for cracked light heads that could cause surgical infections. Affected models include VLT600SF STP with specific serial numbers. Contact for repair or replacement.
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
Maquet is recalling VOLISTA StandOP Surgical Lights because their light head forks might crack during use, potentially causing particles to fall into the surgical field and lead to patient infections. This is a serious safety issue that requires immediate attention to prevent infection risks during surgery.
The coating on the light head forks of these surgical lights may crack, allowing small particles to fall into the surgical field during procedures. If these particles enter the surgical site, they could cause infections in patients. This risk occurs during normal surgical use when the lights are operating.
Surgical teams using the Maquet VOLISTA StandOP Surgical Lights with the specific model and serial numbers listed in the recall are most at risk. Patients undergoing surgery with these lights could face infection if particles fall into the surgical field.
Check if your surgical light has the model VLT600SF STP and serial numbers listed: 28004, 31006, 31065, 31066, 31067, 31068, 31071, 35010, 35012, 35044, 35045. The product code is ARD568811910.
Verify your surgical light model is VLT600SF STP with the listed serial numbers to confirm it is affected.
Reach out to Maquet directly for instructions on repair, replacement, or other remedies as specified in the recall notice.
The coating on the light head forks may crack, causing particles to fall into the surgical field and potentially lead to patient infections during surgery.
The affected model is VLT600SF STP with specific serial numbers: 28004, 31006, 31065, 31066, 31067, 31068, 31071, 35010, 35012, 35044, 35045.
No, but you should check if your light is affected and contact Maquet for instructions on repair or replacement.
We’ve drafted a message you can send Maquet to request your refund or repair — edit it as you like.
Subject: Recall Z-1853-2019 — Maquet Maquet Hello, I own a Maquet that is covered by recall Z-1853-2019 from Maquet. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.