Bard recalls GlidePath Long-Term Hemodialysis Catheters with specific model and lot numbers due to molded tip breaking during placement, potentially causing complications.
The action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
Bard is recalling GlidePath Long-Term Hemodialysis Catheters with model code 5396280 (14.5F, Alphacurve, 28cm) because the molded tip may break during catheter placement. This could lead to complications if the catheter becomes dislodged or damaged.
The molded tip of the catheter can break during placement, which may cause the catheter to become dislodged or damaged. This could lead to complications such as infection or improper function during hemodialysis treatment.
Hemodialysis patients using the recalled catheters are most at risk. The specific model and lot numbers are listed in the recall notice.
Check for the model code 5396280 (14.5F, Alphacurve, 28cm length) on the catheter. The lot number RECW2372 with expiration date 3/31/2020 is also associated with this recall.
Verify the model code (5396280) and lot number (RECW2372) on the catheter.
The molded tip may break during placement, which could cause complications like dislodgement or damage.
Check the model code and lot number on the catheter to see if it's affected. Contact Bard for further instructions.
The recall is due to potential complications from a broken molded tip, but the remedy does not specify immediate stop-use or return.
We’ve drafted a message you can send Bard to request your refund or repair — edit it as you like.
Subject: Recall Z-2389-2019 — Bard Bard Hello, I own a Bard that is covered by recall Z-2389-2019 from Bard. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.