Henry Schein recalls lipid panel test strips with lot P901 due to potential MEMo chip mislabeling, which may cause incorrect analyzer display. Affected users should contact for remedy.
The MEMo chips provided with lot P901 may be mislabeled, the MEMo Chip may display Q901 on the analyzer LCD screen, instead of P901
Henry Schein is recalling lipid panel test strips (Ref Number 570-0414) that may have mislabeled MEMo chips. This could lead to incorrect test results if the analyzer shows Q901 instead of P901. The issue affects users of CardioChek PA or Plus analyzers.
The MEMo chip may display Q901 on the analyzer screen instead of P901, causing incorrect test results. This mislabeling could lead to inaccurate cholesterol measurements, but does not cause physical harm directly.
Users of Henry Schein lipid panel test strips with lot number P901 (expired 8/17/2020) who run tests on CardioChek PA or Plus analyzers are affected. Those who have used the test strips without verifying the MEMo chip label are at higher risk.
Check for lot number P901 on the test strip packaging or the MEMo chip. The test strips expire on 8/17/2020 and are used with CardioChek PA or Plus analyzers.
Verify the MEMo chip displays P901, not Q901, before using the test strips.
No, but you should check the MEMo chip label to ensure it shows P901 before using the test strips.
The analyzer may display incorrect results, so you should not use the test strips if the MEMo chip shows Q901.
Look for lot number P901 on the test strip packaging or the MEMo chip itself.
We’ve drafted a message you can send Henry Schein to request your refund or repair — edit it as you like.
Subject: Recall Z-2158-2019 — Henry Schein Henry Schein Hello, I own a Henry Schein that is covered by recall Z-2158-2019 from Henry Schein. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.