Olympus recalls WM-NP2 and WM-DP2 mobile workstations with nurse control panel arms. If a video monitor weighs over 19.8 pounds, the workstation may become unstable and fall.
WM-NP/DP2 workstations could become unstable and may fall over if the control panel arm is mounted with a video monitor weighing more than 19.8 pounds.
Olymp: Olympus is recalling mobile workstations with nurse control panel arms because they can become unstable and fall over if a video monitor weighs more than 19.8 pounds. This poses a risk of injury from falling.
The workstation could become unstable if a video monitor weighs more than 19.8 pounds. This instability may cause the workstation to fall over, potentially leading to injury.
Users of Olympus WM-NP2 or WM-DP2 mobile workstations with nurse control panel arms are affected. Nurses and healthcare workers who use these workstations with heavy video monitors are most at risk.
Check for Olympus WM-NP2 or WM-DP2 mobile workstations with nurse control panel arms. The product has model number MAJ-1664 and all serial numbers.
Ensure the video monitor weighs 19.8 pounds or less to prevent instability.
The Olympus WM-NP2 or WM-DP2 mobile workstations with nurse control panel arms could become unstable and fall over if a video monitor weighs more than 19.8 pounds.
Check for Olympus WM-NP2 or WM-DP2 mobile workstations with nurse control panel arms. The product has model number MAJ-1664 and all serial numbers.
The recall record does not specify a remedy, so there is no action required beyond checking the weight of the monitor.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-1983-2019 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-1983-2019 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.