Terumo recalls oxygenator model 3CX*NX19RW with possible non-functioning thermistor. Affected units may not monitor blood temperature correctly.
Capiox NX19 Oxygenator may contain a non-functioning arterial thermistor component
Terumo is recalling a specific oxygenator model that may have a faulty thermistor component. This could lead to incorrect blood temperature monitoring during medical use.
The oxygenator may have a thermistor that doesn't work properly. This could cause the device to fail in monitoring blood temperature during medical procedures. Inaccurate temperature readings might lead to complications for patients.
Medical professionals using the Terumo Capiox NX19 Oxygenator with reservoir (West Orientation) model P/N 3CX*NX19RW are most at risk. Patients receiving this device could experience inaccurate blood temperature readings.
Check for the model number 3CX*NX19RW on the oxygenator. Lot numbers XAl 1, XA28, or XC07 indicate affected units. The device has a West Orientation reservoir.
Look for the model number 3CX*NX19RW on the oxygenator.
Reach out to Terumo Cardiovascular Systems Corporation for further assistance.
The oxygenator may have a non-functioning thermistor component, which could lead to inaccurate blood temperature monitoring during medical use.
Check the model number and lot numbers to see if your device is affected. Contact Terumo for further instructions.
Affected units have model number 3CX*NX19RW and lot numbers XAl 1, XA28, or XC07 with West Orientation reservoir.
We’ve drafted a message you can send Terumo to request your refund or repair — edit it as you like.
Subject: Recall Z-2102-2019 — Terumo Terumo Hello, I own a Terumo that is covered by recall Z-2102-2019 from Terumo. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.