Seca recalls baby scale carts with serial numbers 6002142000002-6002142001101 due to risk of wheels detaching, which could cause tipping and serious injury to babies left unattended.
It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.
Seca is recalling baby scale carts with specific serial numbers because their wheels can unexpectedly loosen and detach. If the cart tips over, it could cause a baby to fall and suffer serious injuries if left unattended.
The wheels on these baby scale carts can unexpectedly loosen and detach. If this happens, the cart may become unstable and tip over, causing the scale to fall and potentially leading to a baby falling off the scale and sustaining serious injuries.
Parents or caregivers who own Seca baby scale carts with serial numbers 6002142000002 to 6002142001101 are affected. Babies left unattended in these carts are at highest risk of serious injury.
Check the serial number on the cart, which ranges from 6002142000002 to 6002142001101. These carts were sold with the model numbers 402 and 403.
Immediately stop using the baby scale cart if it has the serial numbers 6002142000002 to 6002142001101.
The affected models are Seca 402 and 403 baby scale carts.
The serial numbers range from 6002142000002 to 6002142001101.
Check the serial number on the cart; if it falls between 6002142000002 and 6002142001101, it is affected.
We’ve drafted a message you can send Seca to request your refund or repair — edit it as you like.
Subject: Recall Z-1429-2020 — Seca Seca Hello, I own a Seca that is covered by recall Z-1429-2020 from Seca. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.